Ortho-Clinical Diagnostics, Inc.: Device Recall

Recall #Z-2122-2025 · 06/10/2025

Class II: Risk

Recall Details

Recall Number
Z-2122-2025
Classification
Class II
Product Type
Device
Recalling Firm
Ortho-Clinical Diagnostics, Inc.
Status
Ongoing
Date Initiated
06/10/2025
Location
Rochester, NY, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
27, 330 US/ 19,679,730 OUS

Reason for Recall

Quality control (QC) drift when certain lots of Performance Verifiers (PVs) are used with certain lots of cholinesterase enzyme (CHE) slides. This drift occurs when the PVs are used 2-7 days after reconstitution; however, the label stability claims for the PV fluids is 7 days following reconstitution. This may lead to inaccurate QC results. The use of the PVs when the QC results are outside of the customer s established ranges could result in delayed results for patients due to the need to repeat QC testing or reconstitute new PV fluids and repeat testing. It could also result in erroneous patient results (small positive bias).

Product Description

VITROS Chemistry Products CHE Slides; Catalog Numbers: (1) 191 4605 (300 slides), (2) 800 4707 (90 slides);

Distribution Pattern

Worldwide Distribution: US (nationwide) to states of: AZ, CA, FL, GA, IL, MD, MI, MO, MS, NY, TN, TX, VA; and OUS (International) to countries of: Australia, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Portugal, Russia, Singapore, Spain, Sweden, and United Kingdom.

Other Recalls by Ortho-Clinical Diagnostics, Inc.

View all recalls by Ortho-Clinical Diagnostics, Inc. →

Data from FDA openFDA enforcement reports. This site is not affiliated with or endorsed by the FDA. Recall information may be updated by the FDA at any time.