Ortho-Clinical Diagnostics, Inc.: Device Recall

Recall #Z-2120-2025 · 06/10/2025

Class II: Risk

Recall Details

Recall Number
Z-2120-2025
Classification
Class II
Product Type
Device
Recalling Firm
Ortho-Clinical Diagnostics, Inc.
Status
Ongoing
Date Initiated
06/10/2025
Location
Rochester, NY, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
37429 units (US)/11,703 units (OUS)

Reason for Recall

Quality control (QC) drift when certain lots of Performance Verifiers (PVs) are used with certain lots of cholinesterase enzyme (CHE) slides. This drift occurs when the PVs are used 2-7 days after reconstitution; however, the label stability claims for the PV fluids is 7 days following reconstitution. This may lead to inaccurate QC results. The use of the PVs when the QC results are outside of the customer s established ranges could result in delayed results for patients due to the need to repeat QC testing or reconstitute new PV fluids and repeat testing. It could also result in erroneous patient results (small positive bias).

Product Description

Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Products Performance Verifier I Model/Catalog Number: 806 7324 Software Version: Not Applicable Product Description: VITROS Chemistry Products Performance Verifier is an assayed control used to monitor performance on VITROS 250/350/5,1 FS/4600/XT 3400 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems. VITROS Performance Verifier is prepared from processed human serum to which enzymes, electrolytes, stabilizers,mpreservatives and other organic analytes have been added. Component: No, not a component

Distribution Pattern

Worldwide Distribution: US (nationwide) to states of: AZ, CA, FL, GA, IL, MD, MI, MO, MS, NY, TN, TX, VA; and OUS (International) to countries of: Australia, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Portugal, Russia, Singapore, Spain, Sweden, and United Kingdom.

Other Recalls by Ortho-Clinical Diagnostics, Inc.

View all recalls by Ortho-Clinical Diagnostics, Inc. →

Data from FDA openFDA enforcement reports. This site is not affiliated with or endorsed by the FDA. Recall information may be updated by the FDA at any time.