Boston Scientific Corporation: Device Recall
Recall #Z-2105-2025 · 06/10/2025
Recall Details
- Recall Number
- Z-2105-2025
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Boston Scientific Corporation
- Status
- Ongoing
- Date Initiated
- 06/10/2025
- Location
- Saint Paul, MN, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 14,724
Reason for Recall
Potential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may go undetected in clinical settings due to small size and variable location.
Product Description
VersaCross Transseptal Sheath Kit (VXS) Product Description/IDs VXS0201 VXS 63CM/45DEG/D0 SHEATH VXS2112 VERSACROSS SHEATH (63CM 45DEG D0)-J Used for the percutaneous introduction of various types of cardiovascular catheters and guidewires to all heart chambers, including the left atrium via transseptal perforation / puncture.
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Canada, Hong Kong, and Japan.