Becton, Dickinson and Company, BD Bio Sciences: Device Recall

Recall #Z-2078-2025 · 06/03/2025

Class II: Risk

Recall Details

Recall Number
Z-2078-2025
Classification
Class II
Product Type
Device
Recalling Firm
Becton, Dickinson and Company, BD Bio Sciences
Status
Ongoing
Date Initiated
06/03/2025
Location
Milpitas, CA, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
590 units

Reason for Recall

Due to customer complaints, errors in spectral overlap factors table for 7-color setup beads may affect accurate setup of the flow cytometry instrument prior to analyzing samples.

Product Description

REF: 335775 , BD FACS, 7-Color Setup Beads, 25 setups per kit, Rx Only, IVD. Used in flow cytometry for fluorescence compensation.

Distribution Pattern

US: Arkansas, Arizona, California, Colorado, Connecticut, District of Columbia, Delaware, Florida, Georgia, Iowa, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maryland, Michigan, Minnesota, North Carolina, Nebraska, New Jersey, Nevada, New York, Ohio, Oklahoma, Oregon, Pennsylvania, Puerto Rico, Rhode Island, South Dakota, Tennessee, Texas, Utah, Virginia, Wisconsin, West Virginia Argentina, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Dominican Republic, Ecuador, Hong Kong, India, Japan, South Korea, Mexico, Malaysia, Nepal, New Zealand, Peru, Pakistan, Singapore, El Salvador, Taiwan OUS:

Other Recalls by Becton, Dickinson and Company, BD Bio Sciences

View all recalls by Becton, Dickinson and Company, BD Bio Sciences →

Data from FDA openFDA enforcement reports. This site is not affiliated with or endorsed by the FDA. Recall information may be updated by the FDA at any time.