Ortho-Clinical Diagnostics, Inc.: Device Recall
Recall #Z-2073-2025 · 06/12/2025
Recall Details
- Recall Number
- Z-2073-2025
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Ortho-Clinical Diagnostics, Inc.
- Status
- Ongoing
- Date Initiated
- 06/12/2025
- Location
- Rochester, NY, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 2,017 units
Reason for Recall
Lots from Generation (GEN) 15 of VITROS Chemistry Products OP Reagent may generate lower than expected quality control (QC) results. If the customer is unable to get passing quality control results, then the customer would be unable to run the OP-LO protocol leading to a potential delay in patient results.
Product Description
VITROS Chemistry Products OP Reagent Gen 15. Model Number: 680 1997. The VITROS OP Reagent is a dual-chambered package containing ready-to-use liquid reagents that are used to detect opiates in urine. One sales unit contains 6 Microtip packs of reagent. Each pack has 50 OP tests.
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, France, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.