Beckman Coulter Inc.: Device Recall

Recall #Z-2028-2025 · 04/08/2025

Class II: Risk

Recall Details

Recall Number
Z-2028-2025
Classification
Class II
Product Type
Device
Recalling Firm
Beckman Coulter Inc.
Status
Ongoing
Date Initiated
04/08/2025
Location
Brea, CA, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1,805 systems

Reason for Recall

Due to a software issue, the software may incorrectly process the date of birth (DoB) received from the Laboratory Information System (LIS) and when the DoB is set to 01/01/1900 or left empty, REMISOL Advance may incorrectly identify a patient as a newborn, affecting the execution of validation rules based on incorrect reference ranges. The test results are not changed by this software defect and are accurately displayed by REMISOL Advance; however, the evaluation flagging of the results may be affected. This issue may lead to wrong reference range calculations and can result in erroneous flagged results being auto validated and uploaded to LIS.

Product Description

REMISOL Advance UDI-DI code: 13700962601874 All Software Versions: B92487; B92488; C24317; C28652; C37500; D04164; C44703; C57017; C69412; C69413; C73942; C73941; C88470; C88471 The Normand Info REMISOL Advance Data Manager collects and manages data and workflow for 128 connections (Beckman Coulter Instruments, automations, LIS&). It can also be connected to a host system. It provides data management capabilities such as automatic results validation, delta checking, reflex testing, quality control, results editing, archiving and restoration of patient results.

Distribution Pattern

Worldwide - U.S. Nationwide distribution including in the states of AK, AL,AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The countries of Australia, Austria, Bangladesh, Brazil, Brunei Darussalam, Bulgaria, Canada, China, Czechia, Denmark, Ethiopia, France, French Polynesia, Germany, Greece, Guam, Hong Kong, India, Ireland, Italy, Japan, Korea, Republic of, Kuwait, Kyrgyzstan, Lebanon, Malaysia, Mexico, Morocco, Netherlands, New Zealand, North Macedonia, Norway, Poland, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Province of China, Turkey, Uganda, United Arab Emirates, United Kingdom of Great Britain and Northern Ireland, and Zimbabwe