Beckman Coulter, Inc.: Device Recall

Recall #Z-1996-2025 · 05/07/2025

Class II: Risk

Recall Details

Recall Number
Z-1996-2025
Classification
Class II
Product Type
Device
Recalling Firm
Beckman Coulter, Inc.
Status
Ongoing
Date Initiated
05/07/2025
Location
Chaska, MN, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
67 US; 178 OUS

Reason for Recall

Analyzer has calibration issues where curves switched from passed to failed due to system errors during aspiration. This anomaly causes erroneous results and delays if unnoticed. RLU values may be misinterpreted as valid measurements by LIS.

Product Description

DxI 9000 Access Immunoassay Analyzer C11137

Distribution Pattern

Worldwide - US Nationwide distribution in the states of AR, CA, FL, GA, HI, ID, IL, IN, KY, LA, ME, MI, NH, NJ, NY, OH, OK, OR, PA, SD, TN, TX, WA, WI and the countries of Algeria, Australia, Austria, Belgium, Brazil, C¿te d'Ivoire, Croatia, Czechia, France, Germany, Hungary, India, Ireland, Israel, Italy, Kazakhstan, Korea, Republic of, Libya, New Zealand, Oman, Poland, Portugal, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, United Kingdom of Great Britain and Northern Ireland.