ICU Medical, Inc.: Device Recall
Recall #Z-1991-2025 · 04/22/2025
Class II: Risk
Recall Details
- Recall Number
- Z-1991-2025
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- ICU Medical, Inc.
- Status
- Ongoing
- Date Initiated
- 04/22/2025
- Location
- Lake Forest, IL, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 214 units
Reason for Recall
Internal testing found that Plum Duo pumps with software version 1.1.1 have a workflow that may result in bypassing the Maximum Dose Limit alert. The Maximum Dose Limit is set in the LifeShield Drug Library Manager (DLM) within the LifeShield Medication Management Safety Software and specifies the highest dose at which the pump can be programmed for weight or BSA (body surface area) based medication rulesets. The limit can be configured for Dose, Loading Dose, and Bolus Dose.
Product Description
ICU Medical Plum Solos, Item number 400011001; infusion pump
Distribution Pattern
Worldwide - US Nationwide distribution in the states of FL, KY, VA, MS, MA, PA, NY, CA, IA and the country of Philippines.