ICU Medical, Inc.: Device Recall

Recall #Z-1990-2025 · 04/22/2025

Class II: Risk

Recall Details

Recall Number
Z-1990-2025
Classification
Class II
Product Type
Device
Recalling Firm
ICU Medical, Inc.
Status
Ongoing
Date Initiated
04/22/2025
Location
Lake Forest, IL, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1701 units

Reason for Recall

Internal testing found that Plum Duo pumps with software version 1.1.1 have a workflow that may result in bypassing the Maximum Dose Limit alert. The Maximum Dose Limit is set in the LifeShield Drug Library Manager (DLM) within the LifeShield Medication Management Safety Software and specifies the highest dose at which the pump can be programmed for weight or BSA (body surface area) based medication rulesets. The limit can be configured for Dose, Loading Dose, and Bolus Dose.

Product Description

ICU Medical Plum Duo, Item numbers: 1) 400020401, 2) 400021001 , and 3) 400021003; infusion pump

Distribution Pattern

Worldwide - US Nationwide distribution in the states of FL, KY, VA, MS, MA, PA, NY, CA, IA and the country of Philippines.