Ohio Medical Corporation: Device Recall
Recall #Z-1980-2025 · 05/08/2025
Class II: Risk
Recall Details
- Recall Number
- Z-1980-2025
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Ohio Medical Corporation
- Status
- Ongoing
- Date Initiated
- 05/08/2025
- Location
- Gurnee, IL, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 36 units
Reason for Recall
Erroneous calibration values, potentially resulting in: as the device ages, oxygen purity will gradually decrease to the point where the device alarms to indicate low oxygen purity will be incorrect.
Product Description
GCE HEALTHCARE Zen-O lite, Portable Oxygen Concentrator, Model RS-00608-X-S
Distribution Pattern
US Nationwide distribution in the states of AR, FL, VA, ND, MS, MN, CA, NV, KY.