Ohio Medical Corporation: Device Recall

Recall #Z-1980-2025 · 05/08/2025

Class II: Risk

Recall Details

Recall Number
Z-1980-2025
Classification
Class II
Product Type
Device
Recalling Firm
Ohio Medical Corporation
Status
Ongoing
Date Initiated
05/08/2025
Location
Gurnee, IL, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
36 units

Reason for Recall

Erroneous calibration values, potentially resulting in: as the device ages, oxygen purity will gradually decrease to the point where the device alarms to indicate low oxygen purity will be incorrect.

Product Description

GCE HEALTHCARE Zen-O lite, Portable Oxygen Concentrator, Model RS-00608-X-S

Distribution Pattern

US Nationwide distribution in the states of AR, FL, VA, ND, MS, MN, CA, NV, KY.