Baxter Healthcare Corporation: Device Recall
Recall #Z-1964-2025 · 05/12/2025
Class II: Risk
Recall Details
- Recall Number
- Z-1964-2025
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Baxter Healthcare Corporation
- Status
- Ongoing
- Date Initiated
- 05/12/2025
- Location
- Deerfield, IL, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 8027 units
Reason for Recall
Baxter has identified a design issue which may cause the pump to detect the presence of a syringe flange when no syringe is loaded. When this issue occurs, the pumps 'user interface becomes stuck on the "Remove Syringe" screen when an infusion is complete, and the syringe is unloaded. A System Error 21502 (Flange Sensor Failure) may occur. Please note this issue only affects a subset of Novum IQ syringe pumps manufactured or repaired after July 4, 2023.
Product Description
Baxter Novum IQ Syringe Pump, product code 40800BAXUS,
Distribution Pattern
US Nationwide distribution.