Baxter Healthcare Corporation: Device Recall

Recall #Z-1964-2025 · 05/12/2025

Class II: Risk

Recall Details

Recall Number
Z-1964-2025
Classification
Class II
Product Type
Device
Recalling Firm
Baxter Healthcare Corporation
Status
Ongoing
Date Initiated
05/12/2025
Location
Deerfield, IL, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
8027 units

Reason for Recall

Baxter has identified a design issue which may cause the pump to detect the presence of a syringe flange when no syringe is loaded. When this issue occurs, the pumps 'user interface becomes stuck on the "Remove Syringe" screen when an infusion is complete, and the syringe is unloaded. A System Error 21502 (Flange Sensor Failure) may occur. Please note this issue only affects a subset of Novum IQ syringe pumps manufactured or repaired after July 4, 2023.

Product Description

Baxter Novum IQ Syringe Pump, product code 40800BAXUS,

Distribution Pattern

US Nationwide distribution.