Baxter Healthcare Corporation: Device Recall
Recall #Z-1962-2025 · 05/09/2025
Class II: Risk
Recall Details
- Recall Number
- Z-1962-2025
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Baxter Healthcare Corporation
- Status
- Ongoing
- Date Initiated
- 05/09/2025
- Location
- Deerfield, IL, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 277 units
Reason for Recall
The replacement rechargeable batteries inserted into the Welch Allyn PocketScope handles have the potential to cause a short circuit in the handle, which can lead to exposure to excessive heat and user burns.
Product Description
2.5 V replacement rechargeable batteries for Hillrom WELCH ALLYN PocketScope handles, Product Code 72600
Distribution Pattern
US Nationwide distribution.