Natus Neurology DBA Excel Tech., Ltd. (XLTEK): Device Recall
Recall #Z-1925-2025 · 05/07/2025
Recall Details
- Recall Number
- Z-1925-2025
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Natus Neurology DBA Excel Tech., Ltd. (XLTEK)
- Status
- Ongoing
- Date Initiated
- 05/07/2025
- Location
- Oakville, Canada
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 3069
Reason for Recall
Electrode placement labels on the Digital Acquisition Box (DAB) were applied backwards. This may ultimately lead to misdiagnosis of seizure location in neonates monitored with this device.
Product Description
Brand Name: OBM00002 OBM DAB (Digital Acquisition Box) Product Name: OBM00002 OBM DAB (Digital Acquisition Box) Model/Catalog Number: OBM00002 OBM DAB (Digital Acquisition Box) Software Version: N/A Product Description: OBM00002 OBM DAB (Digital Acquisition Box) Component: No
Distribution Pattern
Worldwide distribution - US Nationwide including Puerto Rico (PR) and OUS (foreign) countries of: Italy (IT), Estonia (EE), Israel (IL), China (CN), Germany (DE), Saudi Arabia (SA), Switzerland (CH), Chile (CL), Colombia (CO), Netherlands (the) (NL), Japan (JP), United Kingdom of Great Britain and Northern Ireland(the) (GB), Russian Federation (the)(RU), Bolivia (Plurinational State of) (BO), Canada (CA), France (FR), Peru (PE), Guyana (GY), Mexico (MX), Belgium (BE), Thailand (TH), Paraguay (PY), Hong Kong (HK), United Arab Emirates (AE), Austria (AT), India (IN), Indonesia (ID), Kuwait (KW), Qatar (QA), Lithuania (LT), Denmark (DK), Myanmar (MM), Uruguay (UY), Tunisia (TN), Turkey (TR), Korea (the Republic of) (KR), Hungary (HU), Slovenia (SI), Argentina (AR), Norway (NO), Oman (OM), Poland (PL), Finland (FI), Ireland (IE), Australia (AU), Latvia (LV), Ecuador (EC), Singapore (SG), Spain (ES), Serbia (RS), South Africa (ZA), Romania (RO), Algeria (DZ), Portugal (PT), Dominican Republic (the) (DO), Lebanon (LB), Malaysia (MY), Costa Rica (CR), Viet Nam (VN), & Bahrain (BH).
Other Recalls by Natus Neurology DBA Excel Tech., Ltd. (XLTEK)
- Class II: Risk 06/18/2025