Olympus Corporation of the Americas: Device Recall
Recall #Z-1920-2025 · 05/06/2025
Class II: Risk
Recall Details
- Recall Number
- Z-1920-2025
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Olympus Corporation of the Americas
- Status
- Ongoing
- Date Initiated
- 05/06/2025
- Location
- Center Valley, PA, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 201 units (100 US, 101 OUS)
Reason for Recall
Potential for thermal incident including an unexpected occurrence of smoke, sparks, burning, or unsteady flames during clinical use. The fracturing of the fiber is the cause of, or contributes to, these thermal incidents.
Product Description
Brand Name: Powered Laser Surgical Instrument Product Name: SOLTIVE Pro SuperPulsed Laser System Model/Catalog Number: TFL-SLS, TFL-SLS-JA (Japan only); Product Description: SOLTIVE Laser Systems use technology to increase procedural efficiency and flexibility to ureteroscopy, PCNL, and potential soft tissue applications. Component: No
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of AU, CA, CL, DE, HK, JP, SG.
Other Recalls by Olympus Corporation of the Americas
- Class II: Risk 01/07/2026
- Class II: Risk 01/07/2026
- Class II: Risk 01/07/2026
- Class II: Risk 01/07/2026
- Class II: Risk 01/07/2026
Data from FDA openFDA enforcement reports. This site is not affiliated with or endorsed by the FDA. Recall information may be updated by the FDA at any time.