Civco Medical Instruments Co. Inc.: Device Recall
Recall #Z-1892-2025 · 05/01/2025
Class II: Risk
Recall Details
- Recall Number
- Z-1892-2025
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Civco Medical Instruments Co. Inc.
- Status
- Ongoing
- Date Initiated
- 05/01/2025
- Location
- Kalona, IA, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 506 boxes
Reason for Recall
During the attachment of the needle guide to the covered probe, the tight tolerance and fit of the needle guide can cause a pinhole breach in the transducer cover when latching the guide in place.
Product Description
BX2 Needle Guide, Part Number 644-094, tool for performing needle-guided (or catheter) procedures with the use of diagnostic ultrasound transducers
Distribution Pattern
OUS Distribution to countries of: Japan, Taiwan, France, Georgia, Germany and Denmark