Civco Medical Instruments Co. Inc.: Device Recall

Recall #Z-1892-2025 · 05/01/2025

Class II: Risk

Recall Details

Recall Number
Z-1892-2025
Classification
Class II
Product Type
Device
Recalling Firm
Civco Medical Instruments Co. Inc.
Status
Ongoing
Date Initiated
05/01/2025
Location
Kalona, IA, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
506 boxes

Reason for Recall

During the attachment of the needle guide to the covered probe, the tight tolerance and fit of the needle guide can cause a pinhole breach in the transducer cover when latching the guide in place.

Product Description

BX2 Needle Guide, Part Number 644-094, tool for performing needle-guided (or catheter) procedures with the use of diagnostic ultrasound transducers

Distribution Pattern

OUS Distribution to countries of: Japan, Taiwan, France, Georgia, Germany and Denmark

Data from FDA openFDA enforcement reports. This site is not affiliated with or endorsed by the FDA. Recall information may be updated by the FDA at any time.