Boston Scientific Neuromodulation Corporation: Device Recall
Recall #Z-1890-2024 · 04/18/2024
Recall Details
- Recall Number
- Z-1890-2024
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Boston Scientific Neuromodulation Corporation
- Status
- Ongoing
- Date Initiated
- 04/18/2024
- Location
- Valencia, CA, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 14,977
Reason for Recall
Deep Brain Stimulation (DBS) Implantable Pulse Generator (IPG) may experience routine system check during IPG charging, which may cause device reset. The device reset could lead transient loss of stimulation; patients may experience undesired sensations, transient worsening of movement disorder symptoms, which may lead patient to request surgical intervention for replacement or revision.
Product Description
Vercise Genus Deep Brain Stimulation (DBS) Implantable Pulse Generator IPG: R16 IPG KIT, REF: M365DB12160, and R32 IPG KIT, REF: M365DB12320
Distribution Pattern
Worldwide - US Nationwide distribution including in the states of TX, WA, MT, GA, TN, MA, OR, ID, CA, FL, WI, UT, DE, AZ, PA, DC, LA, KY, VA, MN, NC, OH, NY, MI, CO, IN, MD, NV, AL, IL, AR, SC, OK, MO, MS, NJ, ME, Hi, KS, SD, WV, NM, AK, CT, WY, NE, ND and the countries of Canada, Argentina, Chile, Colombia, Brazil, Algeria, Austria, Belgium, Bulgaria, Croatia, Denmark, Finland, France, Germany, Great Britain, Greece, Hungary, Iceland, Ireland, Italy, Jordan, Kazakhstan, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Saudi Arabia, Serbia, Slovakia, Spain, Sweden, Switzerland, Turkey, Ukraine, United Arab Emirates, Australia, Thailand, Singapore, Hong Kong, Korea, India, Japan.
Other Recalls by Boston Scientific Neuromodulation Corporation
- Class II: Risk 07/08/2025
- Class II: Risk 07/08/2025
- Class II: Risk 06/06/2025
- Class II: Risk 07/17/2024
View all recalls by Boston Scientific Neuromodulation Corporation →