Boston Scientific Neuromodulation Corporation: Device Recall

Recall #Z-1890-2024 · 04/18/2024

Class II: Risk

Recall Details

Recall Number
Z-1890-2024
Classification
Class II
Product Type
Device
Recalling Firm
Boston Scientific Neuromodulation Corporation
Status
Ongoing
Date Initiated
04/18/2024
Location
Valencia, CA, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
14,977

Reason for Recall

Deep Brain Stimulation (DBS) Implantable Pulse Generator (IPG) may experience routine system check during IPG charging, which may cause device reset. The device reset could lead transient loss of stimulation; patients may experience undesired sensations, transient worsening of movement disorder symptoms, which may lead patient to request surgical intervention for replacement or revision.

Product Description

Vercise Genus Deep Brain Stimulation (DBS) Implantable Pulse Generator IPG: R16 IPG KIT, REF: M365DB12160, and R32 IPG KIT, REF: M365DB12320

Distribution Pattern

Worldwide - US Nationwide distribution including in the states of TX, WA, MT, GA, TN, MA, OR, ID, CA, FL, WI, UT, DE, AZ, PA, DC, LA, KY, VA, MN, NC, OH, NY, MI, CO, IN, MD, NV, AL, IL, AR, SC, OK, MO, MS, NJ, ME, Hi, KS, SD, WV, NM, AK, CT, WY, NE, ND and the countries of Canada, Argentina, Chile, Colombia, Brazil, Algeria, Austria, Belgium, Bulgaria, Croatia, Denmark, Finland, France, Germany, Great Britain, Greece, Hungary, Iceland, Ireland, Italy, Jordan, Kazakhstan, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Saudi Arabia, Serbia, Slovakia, Spain, Sweden, Switzerland, Turkey, Ukraine, United Arab Emirates, Australia, Thailand, Singapore, Hong Kong, Korea, India, Japan.