Getinge Usa Sales Inc: Device Recall

Recall #Z-1877-2024 · 05/01/2024

Class I: Dangerous

Recall Details

Recall Number
Z-1877-2024
Classification
Class I
Product Type
Device
Recalling Firm
Getinge Usa Sales Inc
Status
Ongoing
Date Initiated
05/01/2024
Location
Wayne, NJ, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
120 units (US)

Reason for Recall

Sevoflurane used with the vaporizer may degrade to hydrogen fluoride, which could present a risk to patients or health care providers through inhalation and/or skin exposure. Potential harms include irritation of respiratory tract, which in worst case may lead to lung edema and/or severe hypocalcemia, blistering, superficial ulceration, and/or hypomagnesemia.

Product Description

Vaporizer Sevoflurane Maquet Filling, Product code/REF number (6682282), Component of: Flow-i C20, Flow-i C30,Flow-i C40, Flow-c, Flow-e.

Distribution Pattern

Domestic: CA, CO, IN, MO, OR, VA, WA. International: Aruba, Australia, Austria, Belgium, Bolivia, Bosnia-Herzegovina, Brazil, Bulgaria, Chile, China, Costa Rica, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iraq, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Lebanon, Libya, Lithuania, Macau, Macedonia, Malaysia, Mexico, Moldavia, Morocco, Mozambique, Namibia, Netherlands, Nicaragua, Norway, Oman, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Vietnam Please note that a list of foreign consignees was not provided by the manufacturer and it is not available at this time.

Other Recalls by Getinge Usa Sales Inc

View all recalls by Getinge Usa Sales Inc →

Data from FDA openFDA enforcement reports. This site is not affiliated with or endorsed by the FDA. Recall information may be updated by the FDA at any time.