Fresenius Kabi USA, LLC: Device Recall
Recall #Z-1869-2025 · 05/12/2025
Class I: Dangerous
Recall Details
- Recall Number
- Z-1869-2025
- Classification
- Class I
- Product Type
- Device
- Recalling Firm
- Fresenius Kabi USA, LLC
- Status
- Ongoing
- Date Initiated
- 05/12/2025
- Location
- North Andover, MA, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 14,280 units (724 cases)
Reason for Recall
Mis-assembly error of Blood Products Administrations Sets where the 200-micron drip chamber filter may be incorrectly positioned and unable to filter out large blood particulates.
Product Description
IVENIX INFUSION SYSTEM LVP Blood Products Administration Set, Dual-Inlet, Low-Sorbing, Y-Site, Mesh Filter, 15 mL. Intravascular Administration Set.
Distribution Pattern
US distribution: CA, CO, DE, GA, IL, MD, MI, MN, MS, NC, NJ, NV, OK, SC, TX, VA, WA & WI.
Other Recalls by Fresenius Kabi USA, LLC
- Class II: Risk 11/21/2025
- Class I: Dangerous 11/14/2025
- Class I: Dangerous 11/06/2025
- Class II: Risk 11/03/2025
- Class I: Dangerous 11/03/2025
Data from FDA openFDA enforcement reports. This site is not affiliated with or endorsed by the FDA. Recall information may be updated by the FDA at any time.