Philips North America Llc: Device Recall
Recall #Z-1861-2024 · 05/03/2024
Class II: Risk
Recall Details
- Recall Number
- Z-1861-2024
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Philips North America Llc
- Status
- Ongoing
- Date Initiated
- 05/03/2024
- Location
- Cambridge, MA, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 3 units
Reason for Recall
The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire
Product Description
SmartPath to dStream for 3.0T Model Number (REF): 782145
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canada, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lebanon, Libya, Lithuania, Martinique, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Sweden, Switzerland, Thailand, United Kingdom.