Getinge Disinfection Ab: Device Recall
Recall #Z-1721-2025 · 04/02/2025
Recall Details
- Recall Number
- Z-1721-2025
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Getinge Disinfection Ab
- Status
- Ongoing
- Date Initiated
- 04/02/2025
- Location
- Vaxjo, N/A, Sweden
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 424 units (79 US, 345 OUS)
Reason for Recall
The frequency inverter is not programmed by supplier according to specification which causes the circulation pump to start earlier than expected, resulting in additional wear and tear, earlier than expected, on the axial seal in the pump housing.
Product Description
Washer Disinfector Aquadis 56; Model Number: 56M
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Bangladesh, Brazil, Canada, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Italy, Japan, Kenya, Korea, Kuwait, Latvia, Lithuania, Malaysia, Mauritania, Mexico, Netherlands, New Caledonia, Nigeria, Norway, Oman, Pakistan, Philippines, Poland, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Spain, Sweden, Switzerland, Syria, Thailand, Turkey, Turkmenistan, United Arab Emirates, United Kingdom, Vietnam, Zambia.