Getinge Disinfection Ab: Device Recall

Recall #Z-1720-2025 · 04/02/2025

Class II: Risk

Recall Details

Recall Number
Z-1720-2025
Classification
Class II
Product Type
Device
Recalling Firm
Getinge Disinfection Ab
Status
Ongoing
Date Initiated
04/02/2025
Location
Vaxjo, N/A, Sweden
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
102 units (8 US, 94 OUS)

Reason for Recall

The frequency inverter is not programmed by supplier according to specification which causes the circulation pump to start earlier than expected, resulting in additional wear and tear, earlier than expected, on the axial seal in the pump housing.

Product Description

Washer Disinfector Aquadis 56; Model Number: 56A

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Bangladesh, Brazil, Canada, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Italy, Japan, Kenya, Korea, Kuwait, Latvia, Lithuania, Malaysia, Mauritania, Mexico, Netherlands, New Caledonia, Nigeria, Norway, Oman, Pakistan, Philippines, Poland, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Spain, Sweden, Switzerland, Syria, Thailand, Turkey, Turkmenistan, United Arab Emirates, United Kingdom, Vietnam, Zambia.