Boston Scientific Corporation: Device Recall

Recall #Z-1689-2025 · 03/24/2025

Class II: Risk

Recall Details

Recall Number
Z-1689-2025
Classification
Class II
Product Type
Device
Recalling Firm
Boston Scientific Corporation
Status
Ongoing
Date Initiated
03/24/2025
Location
Maple Grove, MN, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
67 units

Reason for Recall

Boston Scientific has identified that IceSeedTM CX needles manufactured between April 2024 and February 2025 have been programmed with DEMO settings instead of Commercial settings (see Affected Product Listing). Upon initial connection of an IceSeed CX needle to the Cryoablation System, the user will be prompted to perform the Needle Integrity and Functionality Test (NIT) per standard protocols. The affected needles perform as intended unless they are disconnected and reconnected to any channel after testing. In this case, the Cryoablation System will prompt re-execution of the NIT.

Product Description

IceSeed 1.5 CX NEEDLE OUS, Cryoablation Needle, REF H7493967534170. The needles are intended to convert high-pressure gas to either a very cold Freezing application or to a warm Thawing application.

Distribution Pattern

Nationwide

Other Recalls by Boston Scientific Corporation

View all recalls by Boston Scientific Corporation →

Data from FDA openFDA enforcement reports. This site is not affiliated with or endorsed by the FDA. Recall information may be updated by the FDA at any time.