Bausch & Lomb Surgical, Inc.: Device Recall
Recall #Z-1680-2025 · 03/27/2025
Class I: Dangerous
Recall Details
- Recall Number
- Z-1680-2025
- Classification
- Class I
- Product Type
- Device
- Recalling Firm
- Bausch & Lomb Surgical, Inc.
- Status
- Ongoing
- Date Initiated
- 03/27/2025
- Location
- Clearwater, FL, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 48,237
Reason for Recall
In response to an increased number of reports of toxic anterior segment syndrome (TASS) in conjunction with implantation of enVista Aspire and enVista Envy IOLs, and certain enVista monofocal lenses in the U.S.
Product Description
Intraocular Lens. enVista Envy IOL, All models starting with EN; enVista Envy Toric IOL, All models starting with ETN.
Distribution Pattern
Worldwide distribution. US Nationwide including PR and GU, and the countries of CAN, AUS, AUT, BEL, BGR, IC, HRV, CZE, FRA, DEU, GRC, HUN, ITA, NLD, POL, PRT, ROU, ESP, GBR, NZL, HKG.