Beckman Coulter, Inc.: Device Recall
Recall #Z-1664-2025 · 03/27/2025
Class II: Risk
Recall Details
- Recall Number
- Z-1664-2025
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Beckman Coulter, Inc.
- Status
- Ongoing
- Date Initiated
- 03/27/2025
- Location
- Chaska, MN, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 370 units
Reason for Recall
When DxI 9000 Access Immunoassay Analyzer is connected to a host system (laboratory information system (LIS) or middleware) and has accumulated canceled QC test results that were not sent by LIS, the analyzer may lose communication with the host system. The lost connection interrupts sample processing, which delays reporting patient test results and may subsequently delay patient treatment.
Product Description
Beckman Coulter DxI 9000 Access Immunoassay Analyzer
Distribution Pattern
Worldwide distribution.