Beckman Coulter, Inc.: Device Recall

Recall #Z-1664-2025 · 03/27/2025

Class II: Risk

Recall Details

Recall Number
Z-1664-2025
Classification
Class II
Product Type
Device
Recalling Firm
Beckman Coulter, Inc.
Status
Ongoing
Date Initiated
03/27/2025
Location
Chaska, MN, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
370 units

Reason for Recall

When DxI 9000 Access Immunoassay Analyzer is connected to a host system (laboratory information system (LIS) or middleware) and has accumulated canceled QC test results that were not sent by LIS, the analyzer may lose communication with the host system. The lost connection interrupts sample processing, which delays reporting patient test results and may subsequently delay patient treatment.

Product Description

Beckman Coulter DxI 9000 Access Immunoassay Analyzer

Distribution Pattern

Worldwide distribution.