IHB OPERATIONS B.V.: Device Recall
Recall #Z-1663-2025 · 03/27/2025
Class II: Risk
Recall Details
- Recall Number
- Z-1663-2025
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- IHB OPERATIONS B.V.
- Status
- Ongoing
- Date Initiated
- 03/27/2025
- Location
- Zwolle, N/A, Netherlands
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 112 units
Reason for Recall
The mounting hardware that is supplied with the 4FXS-60 monitor bracket does not provide adequate thread engagement to support the weight of the frame and monitor. This could lead to a failure of the mounting hardware and subsequent separation of the frame from the mounting column.
Product Description
Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount
Distribution Pattern
US Nationwide distribution including in the states of AL, AZ, CA, FL, IA, ID, IL, LA, MA, MD, MI, MN, NC, NH, NJ, NM, NY, PA, TN, TX, UT, VA, WI, WV.