IHB OPERATIONS B.V.: Device Recall

Recall #Z-1663-2025 · 03/27/2025

Class II: Risk

Recall Details

Recall Number
Z-1663-2025
Classification
Class II
Product Type
Device
Recalling Firm
IHB OPERATIONS B.V.
Status
Ongoing
Date Initiated
03/27/2025
Location
Zwolle, N/A, Netherlands
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
112 units

Reason for Recall

The mounting hardware that is supplied with the 4FXS-60 monitor bracket does not provide adequate thread engagement to support the weight of the frame and monitor. This could lead to a failure of the mounting hardware and subsequent separation of the frame from the mounting column.

Product Description

Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount

Distribution Pattern

US Nationwide distribution including in the states of AL, AZ, CA, FL, IA, ID, IL, LA, MA, MD, MI, MN, NC, NH, NJ, NM, NY, PA, TN, TX, UT, VA, WI, WV.