Radiometer Medical ApS: Device Recall
Recall #Z-1659-2025 · 03/21/2025
Recall Details
- Recall Number
- Z-1659-2025
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Radiometer Medical ApS
- Status
- Ongoing
- Date Initiated
- 03/21/2025
- Location
- Bronshoj, N/A, Denmark
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 6275 analyzers
Reason for Recall
ABL90 FLEX and ABL90 FLEX PLUS analyzers running with Radiometer Windows 10 versions RWIN10 1.0 and RWIN10 1.1 may unexpectedly freeze (the analyzer ceases to operate) during a sample measurement. The analyzer software includes a feature (watchdog) that detects a freeze and automatically restarts the software to continue operating the analyzer. However, even if a restart occurs, any sample being processed at that time will be lost, potential to lead to delay in test results.
Product Description
Radiometer ABL90 Series - Radiometer ABL90 FLEX Model Number 393-090 Radiometer ABL90 Series - Radiometer ABL90 FLEX PLUS Model Number 393-092 The Radiometer ABL90 Series is a portable, automated analyzer for use in the laboratory environment, near patients or point-of-care settings.
Distribution Pattern
U.S.: AL, AR, AZ, BM CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY. O.U.S.: Canada, Afghanistan, Albania, Algeria, Argentina, Bahrain, Bangladesh, Belarus, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Chile, Colombia, Cuba, Cyprus, Dominican Republic, Ecuador, Egypt, Eritrea, Ethiopia, Greece, Guatemala, Honduras, Iraq, Iran, Israel, Jamaica, Jordan, Kenya, Kuwait, Lebanon, Libya, Lithuania, Malta, Mauritius, Moldova, Nepal, Nicaragua, Nigeria, Oman, Pakistan, Paraguay, Peru, Philippines, Puerto Rico, Qatar, Romania, Saudi Arabia, Serbia, Slovenia, Sri Lanka, Suriname, Syria, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Uganda, Ukraine, United Arab Emirates, Uruguay, Uzbekistan, Venezuela, Vietnam, Yemen, Zimbabwe, Australia, New Zealand, New Caledonia, Tokelau, Austria, Belgium, Netherlands, China, Croatia, Czech Republic, Slovakia, Denmark, Iceland, Finland, Estonia, France, Germany, Hong Kong, Hungary, India, Brunei, Italy, Japan, Kazakhstan, Malaysia, Mexico, Norway, Poland, Russia, Singapore, South Africa, Cameroon, Ivory Coast, Malawi, South Korea, Spain, Portugal, Sweden, Switzerland, Turkey, United Kingdom, Ireland
Other Recalls by Radiometer Medical ApS
- Class II: Risk 05/31/2024