Thoratec LLC: Device Recall

Recall #Z-1570-2025 · 03/13/2025

Class I: Dangerous

Recall Details

Recall Number
Z-1570-2025
Classification
Class I
Product Type
Device
Recalling Firm
Thoratec LLC
Status
Ongoing
Date Initiated
03/13/2025
Location
Pleasanton, CA, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1661

Reason for Recall

Left Ventricular Assist System Mobile Power Unit (MPU) may have electrical component that may cause MPU not to turn on, unprompted shut down, or turning off/restarting, may cause system controller to indicate Yellow Wrench or No External Power alarms. If MPU losses power, System Controller can support the pump up to 15 minutes, and if not switched to rechargeable batteries, pump may lose power.

Product Description

HeartMate Mobile Power Unit: REF: 107754, NA; 107758, EU; 107758AU, AU; 107758UK, UK; L107758, EU Rental. Used with HeartMate 3 and II Left Ventricular Assist System.

Distribution Pattern

Worldwide - US Nationwide distribution in the states of WA, NE, VA, PA, DC, KY, CA, IL, WI, TX, MA, AR, MS, MI, OH, FL, NC, TN, LA, NJ, UT, NY, DE, CT, GA, IN, AZ, OK, MO, OR, MN, SC, NM, MD, IA, KS, AL, CO, ME and the countries of ALGERIA, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CHILE, CZECH REPUBLIC, FINLAND, FRANCE, GEORGIA, GERMANY, HONG KONG, ISRAEL, ITALY, JAPAN, KUWAIT, LEBANON, LITHUANIA, NETHERLANDS, NEW ZEALAND, NORWAY, OMAN, POLAND, PORTUGAL, ROMANIA, SERBIA, SINGAPORE, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, TUNISIA, TURKEY, UK, UNITED ARAB EMIRATES.

Data from FDA openFDA enforcement reports. This site is not affiliated with or endorsed by the FDA. Recall information may be updated by the FDA at any time.