Baxter Healthcare Corporation: Device Recall
Recall #Z-1560-2025 · 03/24/2025
Class II: Risk
Recall Details
- Recall Number
- Z-1560-2025
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Baxter Healthcare Corporation
- Status
- Ongoing
- Date Initiated
- 03/24/2025
- Location
- Deerfield, IL, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 41,739 units
Reason for Recall
There were customer reports of devices which experienced battery-related fires. The investigation found that the devices in question were utilizing third-party, after-market batteries that are not approved by Baxter.
Product Description
Welch Allyn CP150 Electrocardiograph:
Distribution Pattern
Worldwide - US Nationwide distribution.