Omnicell, Inc.: Device Recall

Recall #Z-1532-2025 · 02/20/2025

Class II: Risk

Recall Details

Recall Number
Z-1532-2025
Classification
Class II
Product Type
Device
Recalling Firm
Omnicell, Inc.
Status
Ongoing
Date Initiated
02/20/2025
Location
Cranberry Township, PA, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3

Reason for Recall

IVX Station Reconstitution Transfer Sets were shipped in advance of the IVX Station upgrade and not intended for clinical use but are engineering devices intended for use during set up and system calibration. The items were not shipped in a validated packing configuration and sterility cannot be ensured.

Product Description

IVX Station Reconstitution Transfer Set. Model Number: 298921215. Packaging: 1 set per White Tyvek Pouch, 5 Tyvek Pouches per shelf box, 10 shelf boxes per case box. IVX Station Reconstitution Transfer Set is used for the reconstitution of drug vials in hospital pharmacies. It is indicated for use with the IVX Station Pharmacy Compounding System.

Distribution Pattern

Distributed in the US to Alabama

Data from FDA openFDA enforcement reports. This site is not affiliated with or endorsed by the FDA. Recall information may be updated by the FDA at any time.