MEDLINE INDUSTRIES, LP - Northfield: Device Recall

Recall #Z-1431-2025 · 02/12/2025

Class II: Risk

Recall Details

Recall Number
Z-1431-2025
Classification
Class II
Product Type
Device
Recalling Firm
MEDLINE INDUSTRIES, LP - Northfield
Status
Ongoing
Date Initiated
02/12/2025
Location
Northfield, IL, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1669 units

Reason for Recall

The female luer fittings on the Pressure Monitoring Lines (PMLs) were manufactured with excess material near the fluid pathway.

Product Description

namic convenience kits labeled as: 1) MTS,4 VALVE,LFT,HRT,KIT-AHS-A*,NW, REF 60010468; 2) MTS,RIGHT HEART,KIT,W/TRANS,ST. *, REF 60010511; 3) MTS,LT,HT,KIT-BRYAN,LGH,M.C.,EAST, REF 60020323; 4) MTO,RIGHT,KIT-BRONSON,M* HOSP PG, REF 60020676; 5) MTS,LEFT HEART,KIT-C*,MEM HLTHCARE, REF 60035977; 6) MTS,LEFT HEART,KIT,J*,M*,C*, REF 60040273; 7) MTS,LEFT HEART,KIT,C*,C*, REF 60100313; 8) MTS,3V,KIT-M*,MEMORIAL, REF 60131208; 9) MTS,3V,KIT-M*,MEMORIAL, REF 60131208; 10) MTO,LEFT HEART,KIT,-,N* HOSPITAL PG, REF 60140101; 11) MTO,LEFT HEART,KIT,-,N* HOSPITAL PG, REF 60140101; 12) MTS,LEFT HEART,MAN,KIT,S*-BALTIMORE, REF 60190947; 13) MTS-LEFT HEART,KIT-W*,H* HOSP PG, REF 60231052; 14) MTO,3-VALVE,KIT-T*,A*-PG, REF 65190182; 15) MTO,LEFT HEART,KIT,-,S* F*,-,PG, REF 65190474; 16) MTS,DR.,MCKAYS,4V,KIT,-,H* HOSP PG, REF 600800516; 17) MTS,EP,KIT,-,S*,TEACHING HOSPITAL PG, REF 601309013; 18) MTS,RHT,HRT-S* HOSP &,MC, REF 601905413; 19) MTS,PEDS,LFT,&,RT,HRT-S* HOSP, REF 601905414; 20) MTO,SCH,BEL,J*,H*-ST-PAUL,-,PG, REF 606604328; 21) LHK,-,MID,3VCOMP,OFF,PS,S10RA, REF 613000122; 22) MTS,LHK,W* HOSPITAL, REF 6010034281; 23) MTS,LHK,W* HOSPITAL, REF 6010034281; 24) MTS,LHK,P*,P* MED CTR, REF 6010171931; 25) MTS,LHK,M* HEALTHCARE,MGMT, REF 6061073001; 26) TVR US B* INTEGRATED SERVICE CENT, REF 601316711A; 27) LHK-S* MEDICAL CENTER,A*, REF 60190776A; 28) LHK,-,S*,AND,W* HOSPITAL, REF 60191118A.

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of The Netherlands, New Zealand, Slovakia