Beckman Coulter, Inc.: Device Recall

Recall #Z-1387-2025 · 02/06/2025

Class II: Risk

Recall Details

Recall Number
Z-1387-2025
Classification
Class II
Product Type
Device
Recalling Firm
Beckman Coulter, Inc.
Status
Ongoing
Date Initiated
02/06/2025
Location
Chaska, MN, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
80 units

Reason for Recall

Beckman Coulter identified that an issue for DxI 9000 tip stocker interface (TSI) and reagent storage interface (RSI) boards with the firmware version 1.1.0.0 can cause sub-system communication failures. This failure could result in a delay in reporting patient results.

Product Description

DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137, with Component PCB, TSI/RSI firmware C29278

Distribution Pattern

Worldwide - US Nationwide distribution in the states of FL, TX, WA.