Beckman Coulter, Inc.: Device Recall
Recall #Z-1386-2025 · 02/05/2025
Class II: Risk
Recall Details
- Recall Number
- Z-1386-2025
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Beckman Coulter, Inc.
- Status
- Ongoing
- Date Initiated
- 02/05/2025
- Location
- Chaska, MN, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 873 units
Reason for Recall
Beckman Coulter identified that an issue for Access 2 Sample Motors (B48016) due to improper alignment, causing the motors to rub against the sample wheelbase plate and index sensor. The issue was traced to a worn tool used in production (at the supplier), affecting Access 2 sample motors. This could lead to a potential delay in reporting patient test results.
Product Description
Access 2 Immunoassay Analyzer, Catalog Number 81600N
Distribution Pattern
Worldwide - US Nationwide distribution.