ETAC A/S: Device Recall
Recall #Z-1311-2026 · 11/17/2025
Recall Details
- Recall Number
- Z-1311-2026
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- ETAC A/S
- Status
- Ongoing
- Date Initiated
- 11/17/2025
- Location
- Gedved, N/A, Denmark
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 25 units
Reason for Recall
A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a result of deformation of the materials involved between the lifting bar and the arm of the device. Such deformation can lead to limited mobility in the joint, which prevents proper functionality of the equipment.
Product Description
Brand Name: Molift Product Name: Molift 4-point sling bars (Accessory to mobile hoists) Model/Catalog Number: 1430116,1430117,1430115 Product Description: The mobile hoist is an assistive device intended for alleviation of, or compensation for, a functional, impairment due to an injury or disability. The device is designed for an individual lacking the ability to stand up and transfer themselves to / from a bed, the floor, a wheelchair, a chair, a toilet or similar due to reduced mobility or physical strength.
Distribution Pattern
Worldwide - US Nationwide distribution in the state of PA and the countries of Australia, Czech Republic, Finland, France, Germany, Israel, Italy, Japan, New Zealand, Norway, Poland, Qatar Spain, Sweden, Switzerland, UK.
Other Recalls by ETAC A/S
- Class II: Risk 11/17/2025
- Class II: Risk 11/17/2025
- Class II: Risk 11/17/2025