ETAC A/S: Device Recall

Recall #Z-1310-2026 · 11/17/2025

Class II: Risk

Recall Details

Recall Number
Z-1310-2026
Classification
Class II
Product Type
Device
Recalling Firm
ETAC A/S
Status
Ongoing
Date Initiated
11/17/2025
Location
Gedved, N/A, Denmark
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
64 units

Reason for Recall

A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a result of deformation of the materials involved between the lifting bar and the arm of the device. Such deformation can lead to limited mobility in the joint, which prevents proper functionality of the equipment.

Product Description

Brand Name: Molift Product Name: Molift 2-point sling bars (Accessory to mobile hoists) Model/Catalog Numbers: (1) 1530101, (2) 1530102 Product Description: The mobile hoist is an assistive device intended for alleviation of, or compensation for, a functional, impairment due to an injury or disability. The device is designed for an individual lacking the ability to stand up and transfer themselves to / from a bed, the floor, a wheelchair, a chair, a toilet or similar due to reduced mobility or physical strength.

Distribution Pattern

Worldwide - US Nationwide distribution in the state of PA and the countries of Australia, Czech Republic, Finland, France, Germany, Israel, Italy, Japan, New Zealand, Norway, Poland, Qatar Spain, Sweden, Switzerland, UK.