Beckman Coulter, Inc.: Device Recall
Recall #Z-1235-2026 · 12/23/2025
Class II: Risk
Recall Details
- Recall Number
- Z-1235-2026
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Beckman Coulter, Inc.
- Status
- Ongoing
- Date Initiated
- 12/23/2025
- Location
- Miami, FL, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 50
Reason for Recall
Sample preparation system designed to automate staining, lysing, cell washing has a software error that causes samples to be dispensed close to or touching tube bottoms, which could dislodge fluid line, result in sample loss, spillage, make instrument inoperable and will lead to unintentional removal of 60-100 microliters from final sample, may lead to incorrect diagnosis or patient management.
Product Description
CellMek SPS Sample Preparation System, REF: C44603, with software
Distribution Pattern
Worldwide - US Nationwide distribution in the states of MO, WA, MA, IL, AL, OR, IL and the countries of AB, BE, ES, FR, NL, ON, DE, AT, VIC, AU, GB, SK, MX, CH.