Beckman Coulter, Inc.: Device Recall

Recall #Z-1235-2026 · 12/23/2025

Class II: Risk

Recall Details

Recall Number
Z-1235-2026
Classification
Class II
Product Type
Device
Recalling Firm
Beckman Coulter, Inc.
Status
Ongoing
Date Initiated
12/23/2025
Location
Miami, FL, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
50

Reason for Recall

Sample preparation system designed to automate staining, lysing, cell washing has a software error that causes samples to be dispensed close to or touching tube bottoms, which could dislodge fluid line, result in sample loss, spillage, make instrument inoperable and will lead to unintentional removal of 60-100 microliters from final sample, may lead to incorrect diagnosis or patient management.

Product Description

CellMek SPS Sample Preparation System, REF: C44603, with software

Distribution Pattern

Worldwide - US Nationwide distribution in the states of MO, WA, MA, IL, AL, OR, IL and the countries of AB, BE, ES, FR, NL, ON, DE, AT, VIC, AU, GB, SK, MX, CH.