Tyber Medical: Device Recall

Recall #Z-1213-2025 · 01/21/2025

Class II: Risk

Recall Details

Recall Number
Z-1213-2025
Classification
Class II
Product Type
Device
Recalling Firm
Tyber Medical
Status
Ongoing
Date Initiated
01/21/2025
Location
Bethlehem, PA, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
214 units

Reason for Recall

Complaints of the locking screw passing through the locking hole intra-operatively during use of the A.L.P.S. mvX Medial Tibia plates and Anatomic Lateral Fibula. r surgical delays have been reported with the use of the devices

Product Description

A.L.P.S. mvX- ANATOMIC LAT FIB PLATE 8H LT: Anatomic Lateral Fibula Plate, 8-hole, Left Intended to bridge or otherwise stabilize bone fragments to facilitate healing. Model/Catalog Number: 770708081

Distribution Pattern

Nationwide

Data from FDA openFDA enforcement reports. This site is not affiliated with or endorsed by the FDA. Recall information may be updated by the FDA at any time.