Augustine Temperature Management, LLC: Device Recall

Recall #Z-1198-2025 · 12/16/2024

Class II: Risk

Recall Details

Recall Number
Z-1198-2025
Classification
Class II
Product Type
Device
Recalling Firm
Augustine Temperature Management, LLC
Status
Ongoing
Date Initiated
12/16/2024
Location
Minnetonka, MN, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1 unit

Reason for Recall

There is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, the connector cable on the mattress can experience the same type of deflection that caused the intermittent connection with the cables connecting to the ESU.

Product Description

Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 127 cm (50"), REF US502, to prevent or treat hypothermia

Distribution Pattern

US Nationwide distribution in the states of TX, AL, NV; AU, NZ.