RAYSEARCH LABORATORIES AB: Device Recall
Recall #Z-1193-2025 · 01/16/2025
Recall Details
- Recall Number
- Z-1193-2025
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- RAYSEARCH LABORATORIES AB
- Status
- Ongoing
- Date Initiated
- 01/16/2025
- Location
- Stockholm, N/A, Sweden
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 0
Reason for Recall
Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs.
Product Description
Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 6.0.0.24 Software Version: RayStation 6
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Belgium ENG, Canada, China, China APactron, Denmark, France, Germany, Hong Kong S.A.R. China, India, Italy, Japan, Korea, Netherlands, Norway, Poland, Singapore, Spain, SwitzerlandENG, Taiwan, Thailand, UK.