NIPRO Technical Services, Inc.: Device Recall
Recall #Z-1185-2026 · 12/12/2025
Class II: Risk
Recall Details
- Recall Number
- Z-1185-2026
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- NIPRO Technical Services, Inc.
- Status
- Ongoing
- Date Initiated
- 12/12/2025
- Location
- Mesa, AZ, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 499 units
Reason for Recall
due to a defect associated with the conductivity values of the solution used to calibrate meters and result in the reporting of incorrect conductivity values for dialysate.
Product Description
Conductivity Standard Solution Catalog/Model Numbers: 10001, 10011, 10021, 10032, 10040, 10040c, and 10041 Conductivity Standard Solutions are a secondary standard solution used for the calibration of conductivity cells together with conductivity meters
Distribution Pattern
U.S. Nationwide distribution in the states of AL, AR, AZ, CA, DE, FL, GA. IL, IN, KS, MI, MO, NC, ND, NJ, NV, NY, OH, OK, OR, PA, SC, TN TX, UT, VA, and WA.
Data from FDA openFDA enforcement reports. This site is not affiliated with or endorsed by the FDA. Recall information may be updated by the FDA at any time.