Siemens Medical Solutions USA, Inc: Device Recall
Recall #Z-1174-2026 · 12/29/2025
Class II: Risk
Recall Details
- Recall Number
- Z-1174-2026
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Siemens Medical Solutions USA, Inc
- Status
- Ongoing
- Date Initiated
- 12/29/2025
- Location
- Malvern, PA, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 40 units
Reason for Recall
Ortho images acquired with preset automatic horizontal flip are not flipped and when acquiring single images, there is the possibility that an image of a previous patient may be processed.
Product Description
Interventional Fluoroscopic X-Ray System: LUMINOS Q.namix R; Siemens Material Number (SMN): 11574002; With Software Version VB10D-SP02
Distribution Pattern
Worldwide - US Nationwide distribution in the states of AK, IL, NY, VA, WI and the countries of Austria, Belgium, Denmark, France, Germany, Greece, Ireland, Italy, Netherlands, Norway, South Africa, Spain, Sweden, Switzerland.