Siemens Medical Solutions USA, Inc: Device Recall

Recall #Z-1174-2026 · 12/29/2025

Class II: Risk

Recall Details

Recall Number
Z-1174-2026
Classification
Class II
Product Type
Device
Recalling Firm
Siemens Medical Solutions USA, Inc
Status
Ongoing
Date Initiated
12/29/2025
Location
Malvern, PA, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
40 units

Reason for Recall

Ortho images acquired with preset automatic horizontal flip are not flipped and when acquiring single images, there is the possibility that an image of a previous patient may be processed.

Product Description

Interventional Fluoroscopic X-Ray System: LUMINOS Q.namix R; Siemens Material Number (SMN): 11574002; With Software Version VB10D-SP02

Distribution Pattern

Worldwide - US Nationwide distribution in the states of AK, IL, NY, VA, WI and the countries of Austria, Belgium, Denmark, France, Germany, Greece, Ireland, Italy, Netherlands, Norway, South Africa, Spain, Sweden, Switzerland.

Data from FDA openFDA enforcement reports. This site is not affiliated with or endorsed by the FDA. Recall information may be updated by the FDA at any time.