Boston Scientific Corporation: Device Recall
Recall #Z-1161-2026 · 12/19/2025
Recall Details
- Recall Number
- Z-1161-2026
- Classification
- Class I
- Product Type
- Device
- Recalling Firm
- Boston Scientific Corporation
- Status
- Ongoing
- Date Initiated
- 12/19/2025
- Location
- Marlborough, MA, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 266 units
Reason for Recall
Increased reports of stent deployment and expansion issues. The most common outcome reasonably foreseeable resulting from difficulty deploying the stent is prolongation of the procedure to exchange the device for a new one. The most serious reasonably foreseeable outcome in cases where the first flange is unable to deploy or expand would be a cascade of events resulting in additional intervention, either endoscopic or surgical, to remove the stent and close the puncture site.
Product Description
HOT AXIOS Stent and Electrocautery-Enhanced Delivery System 6mm x 8mm; Material Number (UPN): M00553520;
Distribution Pattern
Worldwide distribution. US Nationwide including Puerto Rico; Algeria, Argentina, Austria, Belgium, Brazil, Canada, CHILE, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, JAPAN, Jordan, Kuwait, Lebanon, Luxembourg, Malaysia, Mexico, Morocco, Netherlands, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, PERU, Poland, Portugal, Romania, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, UAE, and United Kingdom.
Other Recalls by Boston Scientific Corporation
- Class I: Dangerous 12/19/2025
- Class I: Dangerous 12/19/2025
- Class I: Dangerous 12/19/2025
- Class I: Dangerous 12/19/2025
- Class I: Dangerous 12/19/2025