Beckman Coulter Inc.: Device Recall

Recall #Z-1112-2025 · 01/02/2025

Class II: Risk

Recall Details

Recall Number
Z-1112-2025
Classification
Class II
Product Type
Device
Recalling Firm
Beckman Coulter Inc.
Status
Ongoing
Date Initiated
01/02/2025
Location
Brea, CA, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
6,994 US; 10,067 OUS

Reason for Recall

Identified reagent lots are experiencing a high rate of calibration failures with failure code of "Max Iterate" on Access 2 and UniCel Dxl Immunoassay Systems. Reporting of results may be delayed due to the time required to achieve a passing calibration curve.

Product Description

Access PCT Reagent Pack, Catalog Number C53987, when used on Access 2 and UniCel Dxl Immunoassay Systems

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Austria, Bahrain, Bangladesh, Belgium, Bosnia and Herzegovina, Brazil, Canada, Chile, China, Croatia, Czechia, Denmark, France, Germany, Greece, Hungary, India, Ireland, Italy, Kazakhstan, Korea, Republic of, Malaysia, Mexico, Peru, Philippines, Poland, Puerto Rico, Romania, Slovakia, South Africa, Spain, Switzerland, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom of Great Britain and Northern Ireland, Viet Nam.