Beckman Coulter Inc.: Device Recall
Recall #Z-1112-2025 · 01/02/2025
Recall Details
- Recall Number
- Z-1112-2025
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Beckman Coulter Inc.
- Status
- Ongoing
- Date Initiated
- 01/02/2025
- Location
- Brea, CA, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 6,994 US; 10,067 OUS
Reason for Recall
Identified reagent lots are experiencing a high rate of calibration failures with failure code of "Max Iterate" on Access 2 and UniCel Dxl Immunoassay Systems. Reporting of results may be delayed due to the time required to achieve a passing calibration curve.
Product Description
Access PCT Reagent Pack, Catalog Number C53987, when used on Access 2 and UniCel Dxl Immunoassay Systems
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Austria, Bahrain, Bangladesh, Belgium, Bosnia and Herzegovina, Brazil, Canada, Chile, China, Croatia, Czechia, Denmark, France, Germany, Greece, Hungary, India, Ireland, Italy, Kazakhstan, Korea, Republic of, Malaysia, Mexico, Peru, Philippines, Poland, Puerto Rico, Romania, Slovakia, South Africa, Spain, Switzerland, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom of Great Britain and Northern Ireland, Viet Nam.