Nobel Biocare: Device Recall

Recall #Z-1101-2025 · 01/10/2025

Class II: Risk

Recall Details

Recall Number
Z-1101-2025
Classification
Class II
Product Type
Device
Recalling Firm
Nobel Biocare
Status
Ongoing
Date Initiated
01/10/2025
Location
Goteborg, Sweden
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1,558 units (90 units - U.S.)

Reason for Recall

Due to manufacturing issue (unintended by-product of the injection molding process) a sharp pin that may puncture the blister packaging and breaching the sterile barrier of the finished product.

Product Description

. Nobel Biocare N1 TiUltra TCC RP 4.0x9mm . Nobel Biocare N1 TiUltra TCC RP 4.0x11mm . Nobel Biocare N1 TiUltra TCC RP 4.0x13mm Nobel Biocare N1 TiUltra TCC implants are intended for use as an endosseous dental implant in the maxilla or mandible for anchoring or supporting dental protheses to restore chewing function

Distribution Pattern

U.S. Nationwide distribution in the states of CA, FL, IL, MA, NJ, NY, OH, PA, UT, VA and WA.

Data from FDA openFDA enforcement reports. This site is not affiliated with or endorsed by the FDA. Recall information may be updated by the FDA at any time.