Medtronic Inc: Device Recall
Recall #Z-1098-2026 · 12/04/2025
Recall Details
- Recall Number
- Z-1098-2026
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Medtronic Inc
- Status
- Ongoing
- Date Initiated
- 12/04/2025
- Location
- Plymouth, MN, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 442
Reason for Recall
Stents may be impacted by a Nitinol anomaly which causes a reduced stent diameter when deployed and corresponding reduced radial force. Issue may result in stent migration.
Product Description
Abre Venous Self-Expanding Stent System, Model numbers AB9G14120090 AB9G12100090 AB9G14080090 AB9G12080090 AB9U12060090 AB9U14100090 AB9U14080090 AB9U12080090 AB9U14060090 AB9G14100090 AB9G12060090
Distribution Pattern
Worldwide - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Chile, Colombia, Costa Rica, Denmark, Dominican Republic, Finland, Germany, Jordan, Lebanon, Malaysia, Netherlands, New Zealand, Peru, Poland, Russia, Singapore, Spain, Switzerland, Taiwan, Thailand, United Arab Emirates, United Kingdom, Uzbekistan, Vietnam.