Medline Industries, LP: Device Recall

Recall #Z-1090-2026 · 12/16/2025

Class II: Risk

Recall Details

Recall Number
Z-1090-2026
Classification
Class II
Product Type
Device
Recalling Firm
Medline Industries, LP
Status
Ongoing
Date Initiated
12/16/2025
Location
Northfield, IL, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
12 units

Reason for Recall

Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.

Product Description

Medline medical procedure convenience kits labeled as: KIT ROBOTICS UROLOGY PROSTATE, REF DYKMBNDL116G

Distribution Pattern

US Nationwide distribution.