Heraeus Medical GmbH (Dental Division): Device Recall
Recall #Z-1081-2026 · 12/09/2025
Recall Details
- Recall Number
- Z-1081-2026
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Heraeus Medical GmbH (Dental Division)
- Status
- Ongoing
- Date Initiated
- 12/09/2025
- Location
- Wehrheim, Germany
- Voluntary/Mandated
- Voluntary: Firm initiated
Reason for Recall
increase of complaints regarding the breakage of the ampoules within their device systems. An extensive investigation revealed that multiple factors have an impact on the ampoule breakage. An unfortunate combination of these influence factors can lead to unbreakable ampoules. If none or not all ampoules break, the product cannot be used on the patient since no or no appropriate cement dough is formed.
Product Description
Brand Name: PALACOS Product Name: PALACOS R+G pro 40; PALACOS R+G pro 80 Model/Catalog Number: 5081288; 5081289 Software Version: N/A Product Description: Radiopaque, poly(methyl methacrylate)-based (PMMA) bone cement, pre-filled into a mixing and application system, suitable for use with or without vacuum (ready to mix). Component: N/A
Distribution Pattern
International distribution in the countries of Aland Islands, Australia, Belgium, Denmark, Finland, Germany, Ireland, Italy, Liechtenstein, Netherlands, New Zealand, Norway, Portugal, Spain, Sweden, Switzerland, United Kingdom.