PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.: Device Recall
Recall #Z-1079-2026 · 12/08/2025
Recall Details
- Recall Number
- Z-1079-2026
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Status
- Ongoing
- Date Initiated
- 12/08/2025
- Location
- Best, Netherlands
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 990 units
Reason for Recall
Nine (9) identified software issues which may result in the following: loss of X-ray after Pedal Tap, loss of X-ray after Phase Fault, missing live X-ray on the FlexVision monitor, system continues to restart, AMD triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after start-up, misalignment of Marker Tool Overlay, longitudinal position error.
Product Description
Philips Azurion Systems R3.0. Includes the following label descriptions and corresponding models: 1. Azurion 3 M12. Model Number: 722229. 2. Azurion 3 M15. Model Number: 722230. 3. Azurion 5 M12. Model Number: 722231. 4. Azurion 5 M20. Model Number: 722232. 5. Azurion 7 B12. Model Number: 722235. 6. Azurion 7 B20. Model Number: 722236. 7. Azurion 7 M12. Model Number: 722233. 8. Azurion 7 M20. Model Number: 722234.
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Brunei Darussal, Bulgaria, Cambodia, Canada, Chile, Colombia, Cote de'Ivore, Croatia, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Honduras, Hong Kong, India, Indonesia, Iraq, Israel, Italy, Japan, Kazakhstan, Kosovo, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Mauritius, Morocco, Nepal, Netherlands, New Zealand, Nicaragua, Norway, Oman, Palestine, Philippines, Poland, Portugal, R¿union, Romania, Saudi Arabia, Serbia, Singapore, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkiye, Ukraine, United Kingdom, Utd.Arab.Emir., Uzbekistan, Vietnam, Yemen.