PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.: Device Recall

Recall #Z-1079-2026 · 12/08/2025

Class II: Risk

Recall Details

Recall Number
Z-1079-2026
Classification
Class II
Product Type
Device
Recalling Firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Status
Ongoing
Date Initiated
12/08/2025
Location
Best, Netherlands
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
990 units

Reason for Recall

Nine (9) identified software issues which may result in the following: loss of X-ray after Pedal Tap, loss of X-ray after Phase Fault, missing live X-ray on the FlexVision monitor, system continues to restart, AMD triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after start-up, misalignment of Marker Tool Overlay, longitudinal position error.

Product Description

Philips Azurion Systems R3.0. Includes the following label descriptions and corresponding models: 1. Azurion 3 M12. Model Number: 722229. 2. Azurion 3 M15. Model Number: 722230. 3. Azurion 5 M12. Model Number: 722231. 4. Azurion 5 M20. Model Number: 722232. 5. Azurion 7 B12. Model Number: 722235. 6. Azurion 7 B20. Model Number: 722236. 7. Azurion 7 M12. Model Number: 722233. 8. Azurion 7 M20. Model Number: 722234.

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Brunei Darussal, Bulgaria, Cambodia, Canada, Chile, Colombia, Cote de'Ivore, Croatia, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Honduras, Hong Kong, India, Indonesia, Iraq, Israel, Italy, Japan, Kazakhstan, Kosovo, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Mauritius, Morocco, Nepal, Netherlands, New Zealand, Nicaragua, Norway, Oman, Palestine, Philippines, Poland, Portugal, R¿union, Romania, Saudi Arabia, Serbia, Singapore, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkiye, Ukraine, United Kingdom, Utd.Arab.Emir., Uzbekistan, Vietnam, Yemen.