PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.: Device Recall
Recall #Z-1069-2026 · 12/15/2025
Recall Details
- Recall Number
- Z-1069-2026
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Status
- Ongoing
- Date Initiated
- 12/15/2025
- Location
- Best, Netherlands
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 21 units
Reason for Recall
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
Product Description
Allura Xper FD10/10; Model Number: 722011;
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Cayman Islands, Chile, China, Colombia, Costa Rica, Cuba, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, France, Germany, Greece, Guatemala, Honduras, India, Italy, Japan, Kenya, Korea, Republic of, Malaysia, Mexico, Nepal, Netherlands, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, South Africa, Spain, Sri Lanka, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela, Yemen.
Other Recalls by PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Class II: Risk 12/15/2025
- Class II: Risk 12/15/2025
- Class II: Risk 12/15/2025
- Class II: Risk 12/15/2025
- Class II: Risk 12/15/2025
View all recalls by PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. →