ROi CPS LLC: Device Recall
Recall #Z-1047-2025 · 12/26/2024
Class I: Dangerous
Recall Details
- Recall Number
- Z-1047-2025
- Classification
- Class I
- Product Type
- Device
- Recalling Firm
- ROi CPS LLC
- Status
- Ongoing
- Date Initiated
- 12/26/2024
- Location
- Republic, MO, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 414 kits
Reason for Recall
There is a potential for the Inline Controller to be using the spring out of specification causing a loss of positive pressure resulting in the Neo-Tee Resuscitator and/or Circuit will not reach the high and low ends of the PIP and PEEP pressure ranges impacting effective ventilation of the patient..
Product Description
regard Item Number: 830106006, LD01114F - Newborn Kit containing the component Neo-Tee Resuscitator. Convenience kit to be used in Labor and Delivery.
Distribution Pattern
Distribution was made to MO. There was no government/military/foreign distribution.