ROi CPS LLC: Device Recall

Recall #Z-1047-2025 · 12/26/2024

Class I: Dangerous

Recall Details

Recall Number
Z-1047-2025
Classification
Class I
Product Type
Device
Recalling Firm
ROi CPS LLC
Status
Ongoing
Date Initiated
12/26/2024
Location
Republic, MO, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
414 kits

Reason for Recall

There is a potential for the Inline Controller to be using the spring out of specification causing a loss of positive pressure resulting in the Neo-Tee Resuscitator and/or Circuit will not reach the high and low ends of the PIP and PEEP pressure ranges impacting effective ventilation of the patient..

Product Description

regard Item Number: 830106006, LD01114F - Newborn Kit containing the component Neo-Tee Resuscitator. Convenience kit to be used in Labor and Delivery.

Distribution Pattern

Distribution was made to MO. There was no government/military/foreign distribution.