GE Medical Systems, LLC: Device Recall
Recall #Z-1043-2025 · 01/28/2025
Recall Details
- Recall Number
- Z-1043-2025
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- GE Medical Systems, LLC
- Status
- Ongoing
- Date Initiated
- 01/28/2025
- Location
- Waukesha, WI, United States
- Voluntary/Mandated
- FDA Mandated
- Product Quantity
- 3149 (1697 US; 1452 OUS)
Reason for Recall
The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure techniques. In all cases, a message still appears on the screen when the AEC limit is reached.
Product Description
Definium 656HD & Discovery XR656HD (GTINs: 00840682138598, 00195278840226, 00840682136709). X-Ray imaging system.
Distribution Pattern
Worldwide Distribution: US (nationwide) including Puerto Rico and OUS (foreign) countries to: Canada, Chile, China, Colombia, Coast Rica, Croatia, Cyprus, Czech Republic, Czechia, Dominican Republic, Egypt, Estonia, Finland, France, Germany, Gibraltar, Greece, Honduras, Hong Kong, Hungary, Iceland, Indonesia, Iraq, Ireland, Italy, Jamaica, Japan, Jordan, Kenya, Korea (Republic of) Kuwait, Libya, Lithuania, Macedonia, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Philippines, Poland, Portugal, Qatar, Republic of Serbia, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, and Vietnam.
Other Recalls by GE Medical Systems, LLC
- Class II: Risk 10/24/2025
- Class II: Risk 09/18/2025
- Class II: Risk 09/18/2025
- Class II: Risk 09/18/2025
- Class II: Risk 08/22/2025